Description
The Model 3050 Moisture Analyzer series measures trace levels of moisture in a gas through the use of a quartz-crystal oscillator sample cell. AMETEK is the leader in quartz-crystal technology, which for thirty years has offered significant advantages over other measurement techniques:
- It is the most accurate trace moisture measurement technology available.
- It responds far faster to both increasing and decreasing moisture levels.
- It is specific to moisture in most applications.
- It provides a much more rugged sensor.
Because of these advantages, the quartz-crystal oscillator has become the industry standard for applications ranging from ultrahigh purity semiconductor gases to natural gas streams containing 30% H2S. Now, the Model 3050 analyzers bring the benefits of quartz-crystal technology to a broad spectrum of moisture applications.
Specifications –
See individual Specification Sheets on the Documentation tab for other Model 3050 versions.
Technology | Quartz-crystal microbalance |
Range | Calibrated from 0.1 to 2500 ppmv. Readout capability in ppmw, lb/mmscf, mg/Nm3 and dew point in ºC or ºF (requires process pressure as an input) |
Reference Gas | Continuously produced using actual sample gas |
Online Verification | Internal moisture source with NIST-traceable calibration enables on-demand verification of analyzer accuracy and responsiveness without uninstalling the analyzer |
Accuracy | ±10% of reading from 1 to 2500 ppmv with standard calibration; special calibration ranges not required |
Reproducibility | ±5.0% of reading from 1 to 2500 ppmv |
Limits of Detection | 0.1 ppmv |
A/D Resolution | 16-bit (0.0015%) |
QCM Response Time | Near real time. Computer-enhanced response, which may lead to errors is not required to obtain quick wet-up or dry-down response |
Sensitivity | 0.1 ppmv or 1% of reading, whichever is greater |
Allowable Inlet Pressure | 1.3 to 3.3 Bar (20 to 50 psi) gauge; up to 200 Bar (3000 psi) with optional pressure reducer; analyzer performance is independent of process pressure |
Exhaust Pressure | 0 to 1 Bar (0 to 15 psi) gauge |
Sample Gas Temperature | 0 to 100ºC (32º to 212ºF); analyzer performance is immune to changes in sample gas |
Gas Flow Requirements | Approx.150 sccm. Approx. 1.0 slpm bypass flow for increased speed of response |
Outputs | Isolated 4-to-20 mA analog signal, 12 bit (0.025%) resolution, RS232 and RS484 serial communication ports (supports Modbus RTU) |
Alarms | Three contact closures; system, data valid, and concentration alarms |
Ambient Temperature Limits | Analyzer: 5º to 50ºC (41º to 122ºF) Enclosed Analyzer with sample system: -20º to 45ºC (-4º to 113ºF) |
Voltage Power Requirements | Analyzer: 24 VDC, 50W Analyzer with sample system: 120 ± 10% VAC, 50/60 Hz, 150W maximum 230 ± 10% VAC, 50/60 Hz, 150W maximum |