Dycor LC-D Series Residual Gas Analyzer

The Dycor LC-D Series Residual Gas Analyzer (RGA) offers outstanding performance.

Description

The Dycor LC-D Series Residual Gas Analyzer (RGA) offers outstanding performance, yet is priced to be the most cost-effective RGA on the market. The LC-D Series instruments are constructed from materials that are compatible with Ultra High Vacuum (UHV) systems and are designed to require the very minimum in maintenance. This versatile RGA is ideally suited for use with UHV systems and equipment monitoring of semiconductor and thin film processes in the high vacuum range. Coupled with the System 2000 software (which is 32-bit, multi-threaded, and designed for use with Windows 98 through Windows 8), the LC-D provides advanced features that are associated with higher priced instruments.

The LC-D Series RGA includes Dycor’s 1-100 AMU-range open ion source analyzer head with dual filaments and Faraday cup detector, electrometer preamplifier, and RF power supply with an Ethernet port for interfacing to a PC. Options include 1-200 AMU range and 1-300 AMU range, and channel plate electron multiplier.

  • Alarms – Every data channel has low, low warning, high warning, and high alarms.
  • Auto Tune – Automatically sets RF-tune, mass position and resolution.
  • Printing – Print one, some, or all data displays.
  • Ethernet Connectivity – 10/100BASE-T Ethernet – Auto-Sensing.

Software Features

  • Ethernet Connectivity: 10/100BASE-T Ethernet – Auto-Sensing
  • Variety of operating modes including Analog and Bar, Meter, Annunciator, Tabular, Trend, and Leak Detection
  • Custom modes to combine display windows from other modes
  • Data that can be viewed in real time or saved to a file for later retrieval and comparison with current spectra
  • Auto-Tune capabilities to ensure high quality, repeatable data
  • Library Mode for comparison between collected sample spectra and stored library spectra. Facilitated by split-screen display
  • OPC, MODBUS & Dynamic Data Exchange (DDE) automation
  • Alarms: Low, Low Warning, High Warning, High

Specifications

Mass Range 1 – 100 AMU standard; 1 – 200 and 1 – 300 AMU optional
Operating Pressure Range 10 -4 Torr to ultrahigh vacuum
Minimum Detectable Partial Pressure 5 x 10 -12 Torr (5 x 10 -14 Torr for electron multiplier units)
Resolution Adjustable to constant peak width (0.5 AMU at 10% height)
Emission Current 0.1 to 10 mA; 50 mA to degas
Electron Energy 30 to 150 volts to operate; 200 volts to degas
Ion Energy 1 to 10 volts
Source Sensitivity (Faraday Cup) 2 x 10 -4 amps per Torr at detector (measured with nitrogen at mass 28) with peak width = 0.5 at 10% height and 1 x 10 -3 amps emission current
Power Requirements 24 VDC @ 3 Amps; 110/220 VAC, 50/60 Hz adaptor available
Stability
  Mass Stability ± 0.1 AMU after 30- minute warm-up
  Peak Height ± 2% after 30-minute warm-up
Minimum PC Requirements Windows 98 through Windows 8
Physical Dimensions
  Weight 4.6 lb. (2.1 kg) without head
7.4 lb. (3.4 kg) with head
  Width 4.5 in. (11.4 cm)
  Length 9 in. (22.9 cm)
  Height 5.25 in. (13.3 cm)

Attachements

LC-D Series Residual Gas Analyzer Productsheet