Model 5920 UHP Moisture Analyzer

AMETEK’s Model 5920 is an easy-to-use process moisture analyzer that offers a truly remarkable combination of performance features: multi-gas compatibility, exceptional accuracy, on-line verification, fast response speed, and wide measurement range.

Description

Building upon the success of our quartz-crystal moisture measurement technology, AMETEK’s Model 5920 is an easy-to-use process moisture analyzer that offers a truly remarkable combination of performance features: multi-gas compatibility, exceptional accuracy, on-line verification, fast response speed, and wide measurement range. The 5920 is ideal for critical moisture measurement applications including high purity gas production and semiconductor gases. Equipped with an on-line verification system, this analyzer is designed to rapidly build and maintain operator confidence in its analyses. The verification system allows you to challenge the analyzer’s sensor at will using a zero gas to check its baseline stability to confirm the sensor’s stability and sensitivity to low ppbv moisture concentrations.

Specifications

Compatible Gases Inerts (He, Ar, Ne, Xe, Kr) O2, H2, N2. Contact AMETEK to confirm compatibility with other gases.
Range Calibrated from 0 to 150 ppbv. Trend indication to 1000 ppbv.
Limit of Detection 1.0  ppbv nominal
Accuracy ±1 ppbv or ±10% of the reading, whichever is greater
Sensitivity Better than 1 ppbv
Response Time Typically 80% of a 25 ppbv step change in either direction in 10 minutes or less
Inlet Pressure 138 to 345 kPa (20 to 50 psig). Specified performance is obtained when the inlet gas pressure is maintained within ±17 kPa (±2.5 psi)
Exhaust Pressure Atmospheric
Sample Flow Requirement <0.5 slpm
Inlet Gas Temperature 0º to 100ºC (32º to 212ºF). Optimal results are obtained when the inlet gas temperature is maintained at 60ºC (140ºF)
Outputs Four line by 20 character LCD display
One self-powered 4 to 20 mA, into 100 to 500 ohm load analog output; can be configured for loop-powered operation
RS485 and RS232 serial ports
Alarms System alarm, concentration alarm, and data valid 30 VAC or 60 VDC max, 50 VA or 1A max, resistive
Environmental Conditions Ambient temperature range 10º to 30ºC (50º to 86ºF)
Optimal results are obtained when ambient is maintained within ±5ºC (±9ºF)
Relative humidity 90%, noncondensing
Pollution degree 2
Maximum altitude 2000 meters (6560 ft)
Installation category II
Indoor Use Only
Utility Requirements 100 to 132 VAC or 230 VAC ±10%, 47 to 63 Hz, 185W
Instrument Air: 550 to 690 kPa (80 to 100 psi)
-40ºC dew point
Mounting Configuration 19″ rack
Dimensions(WxHxD) 48×17.7×50.9 cm (19x7x20 in)
Net Weight 15.9 kg (35 lb)

Attachements

Model 5920 Moisture Analyzer Productsheet